The methods of determination of Be, Cu, Fe, Ni, Si, Ti, and Cd in certified standard samples of magnesium alloys by inductively coupled plasma mass spectrometry (ICP-MS) are discussed. The procedure for dissolving the sample and preparing it for analysis is given. To eliminate spectral interference, a collision-reaction cell was used. An internal standard is selected and the accuracy characteristics of the magnesium alloy analysis method are calculated. The accuracy of the results is confirmed by analyzing a certified standard sample of magnesium alloy.