Determination of Impurities in Magnesium Alloys by Inductively Coupled Plasma Mass Spectrometry
- 作者: Alekseev A.V.1, Yakimovich P.V.1
-
隶属关系:
- State Research Center of the Russian Federation – All-Russia Research Institute of Aviation Materials (VIAM)
- 期: 卷 62, 编号 8 (2019)
- 页面: 741-745
- 栏目: Physicochemical Measurements
- URL: https://bakhtiniada.ru/0543-1972/article/view/246772
- DOI: https://doi.org/10.1007/s11018-019-01688-w
- ID: 246772
如何引用文章
详细
The methods of determination of Be, Cu, Fe, Ni, Si, Ti, and Cd in certified standard samples of magnesium alloys by inductively coupled plasma mass spectrometry (ICP-MS) are discussed. The procedure for dissolving the sample and preparing it for analysis is given. To eliminate spectral interference, a collision-reaction cell was used. An internal standard is selected and the accuracy characteristics of the magnesium alloy analysis method are calculated. The accuracy of the results is confirmed by analyzing a certified standard sample of magnesium alloy.
作者简介
A. Alekseev
State Research Center of the Russian Federation – All-Russia Research Institute of Aviation Materials (VIAM)
编辑信件的主要联系方式.
Email: kvark-87@mail.ru
俄罗斯联邦, Moscow
P. Yakimovich
State Research Center of the Russian Federation – All-Russia Research Institute of Aviation Materials (VIAM)
Email: kvark-87@mail.ru
俄罗斯联邦, Moscow
补充文件