An automatic digital tensimeter with a membrane zero-manometer
- 作者: Doinikov D.A.1, Kazakov I.V.1, Krasnova I.S.1, Timoshkin A.Y.1
-
隶属关系:
- Institute of Chemistry
- 期: 卷 91, 编号 8 (2017)
- 页面: 1603-1608
- 栏目: Physicochemical Research Methods
- URL: https://bakhtiniada.ru/0036-0244/article/view/169680
- DOI: https://doi.org/10.1134/S0036024417080088
- ID: 169680
如何引用文章
详细
An automatic setup for measuring vapor pressure and its temperature dependence in a closed volume is designed. The setup can be used to investigate chemical equilibria in both condensed phase–vapor and homogeneous gas-phase systems, along with the kinetics of irreversible processes with the evolution of gas substances. The digital tensimeter offers an improved version of static tensimetry with a membrane zeromanometer. In contrast to conventional optical methods for recording the position of a zero-manometer, an approach based on the Hall effect is used. Automation of the experiment and data acquisition greatly increases the number of experimental points and reduces the errors related to measurements. With processes of irreversible hydrogen release, the method detects hydrogen with a precision of 2 × 10−6 g, remarkably exceeding the possibilities of thermogravimetric methods.
作者简介
D. Doinikov
Institute of Chemistry
编辑信件的主要联系方式.
Email: d.a.doinikov@gmail.com
俄罗斯联邦, St. Petersburg, 198504
I. Kazakov
Institute of Chemistry
Email: d.a.doinikov@gmail.com
俄罗斯联邦, St. Petersburg, 198504
I. Krasnova
Institute of Chemistry
Email: d.a.doinikov@gmail.com
俄罗斯联邦, St. Petersburg, 198504
A. Timoshkin
Institute of Chemistry
Email: d.a.doinikov@gmail.com
俄罗斯联邦, St. Petersburg, 198504
补充文件