Clinical case of intracardiac pacemaker lead fracture

Cover Page

Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

Lead fracture is a serious complication of a pacemaker; its prevalence is 0.1–4.2%. The common site of lead fracture is at the space between the clavicle and the first rib. The causes are intense physical activity, which approximates the clavicle to the rib and compression of the lead, chest trauma, anatomical features, twiddler’s syndrome. Diagnosis of lead fracture is can be made by electrocardiography — there is a transient or permanent stimulation/sensitivity disturbance. When the programmer interrogates the pacemaker, a significant sign is an abrupt rise in the lead impedance, although cases of fracture with normal impedance values have been reported. The article presents an extremely rare clinical case of an intracardiac lead fracture in a 28-year-old patient. At the initial implantation, leads were passed through the accessory left superior vena cava, resulting in a loop in the right ventricle. The patient himself was subjected to increased physical activity. The question of the need to remove such leads remains open. Some authors note that the distal end is firmly fixed to the heart wall, and therefore does not expose the patient to a vital risk. Others consider that the lead can become a source of thrombus formation, or fragmentation with embolism in the pulmonary circulation can occur. In our case, the causes of the fracture were probably an intense physical activity and bending of the lead inside the right ventricle. The clinical situation was discussed with cardiac surgeons of the federal centers of cardiovascular surgery. Given the high risks of open-heart surgery, it was deci¬ded to refrain from removing the broken lead, and the patient was provided with atrial pacing.

About the authors

R E Kalinin

Ryazan State Medical University

Email: povarov.vladislav@mail.ru
Russian Federation, Ryazan, Russia

I A Suchkov

Ryazan State Medical University

Email: povarov.vladislav@mail.ru
Russian Federation, Ryazan, Russia

V O Povarov

Ryazan State Medical University; Ryazan State Cardiologic Dispensary

Author for correspondence.
Email: povarov.vladislav@mail.ru
Russian Federation, Ryazan, Russia; Ryazan, Russia

I I Shitov

Ryazan State Cardiologic Dispensary

Email: povarov.vladislav@mail.ru
Russian Federation, Ryazan, Russia

S S Potehinskiy

Ryazan State Cardiologic Dispensary

Email: povarov.vladislav@mail.ru
Russian Federation, Ryazan, Russia

D V Solovov

Ryazan State Cardiologic Dispensary

Email: povarov.vladislav@mail.ru
Russian Federation, Ryazan, Russia

S A Peshkov

Ryazan State Cardiologic Dispensary

Email: povarov.vladislav@mail.ru
Russian Federation, Ryazan, Russia

References

  1. Kleemann T., Becker T., Doenges K. et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation. 2007; 115: 2474–2480. doi: 10.1161/CIRCULATIONAHA.106.663807.
  2. Bohm A., Duray G., Kiss R.G. Traumatic pacema­ker lead fracture. Emerg. Med. J. 2013; 30 (8): 686. doi: 10.1136/emermed-2012-202090.
  3. Khattak F., Khalid M., Gaddam S. et al. A rare case of complete fragmentation of pacemaker lead after a high-velocity theme park ride. Case Rep. Cardiol. 2018; 2018: 4192964. doi: 10.1155/2018/4192964.
  4. Böhm A., Komáromy K., Pintér A., Préda I. Pacemaker lead fracture due to Twiddler's syndrome. Pacing Clin. Electrophysiol. 1998; 21 (5): 1162–1163. doi: 10.1111/j.1540-8159.1998.tb00166.x.
  5. De Maria E., Fontana P.L., Bonetti L., Cappelli S. An unusual presentation of complete pacing lead fracture with very low impedance values. J. Cardiovasc. Electrophysiol. 2013; 24 (12): 1225–1425. doi: 10.1111/jce.12223.
  6. Kalinin R.E., Such­kov I.A., Mzhavanadze N.D., Povarov V.O. Endothelial dysfunction in patients with cardiac implantable electro­nic devices (literature review). Nauka molodykh — Eruditio Juvenium. 2016; (3): 84–92. (In Russ.)
  7. Alt E., Volker R., Blomer H. Lead fracture in pacemaker patients. Thorac. Cardiovasc. Surg. 1987; 35: 101–104. doi: 10.1055/s-2007-1020206.
  8. Godin B., Savoure A., Bauer F., Anselme F. Complete pacemaker lead fracture potentially due to intra-cardiac mass. Europace. 2011; 13 (4): 593–595. doi: 10.1093/europace/euq406.
  9. Nägele H., Hashagen S., Ergin M. et al. Coronary sinus lead fragmentation 2 years after implantation with a retained guidewire. Pacing Clin. Electrophysiol. 2007; 30 (3): 438–439. doi: 10.1111/j.1540-8159.2007.00688.x.
  10. Augustine D.X., Carson K., Garg A. Distal pacemaker lead fracture: a rare entity. BMJ Case Rep. 2010; 2010: bcr0520103019. doi: 10.1136/bcr.05.2010.3019.
  11. Mucha E., Catalano P., Myers T. Fracture and migration of a pacemaker atrial lead retention wire found by fluoroscopic screening in an asymptomatic patient. South. Med. J. 1996; 89 (8): 798–800. doi: 10.1097/00007611-199608000-00008.
  12. Kalinin R.E., Suchkov I.A., Shitov I.I. et al. Venous thromboembolic complications in patients with cardiovascular implantable electronic devices. Angiologiya i sosudistaya khirurgiya. 2017; 23 (4): 69–74. (In Russ.)
  13. Golukhova E.Z., Revishvili A.Sh., Bazaev V.A. et al. Dislocation of ventricular electrode of pacemaker into the right pulmonary vein. Vestnik aritmologii. 2012; (67): 66–71. (In Russ.)
  14. Udyavar A.R., Pandurangi U.M., Latchumanadhas K., Mullasari A.S. Repeated fracture of pacemaker leads with migration into the pulmonary circulation and temporary pacemaker wire insertion via the azygous vein. J. Postgrad Med. 2008; 54 (1): 28–31. doi: 10.4103/0022-3859.39187.
  15. Kalinin R.E., Suchkov I.A., Mzhavanadze N.D. et al. Congenital complete heart block in pregnant women: a multidisciplinary approach to diagnostics and treatment. I.P. Pavlov Russian Medical Biological Herald. 2016; 24 (3): 79–85. (In Russ.) doi: 10.17816/PAVLOVJ2016379-85.
  16. Kusumoto F.M., Schoenfeld M.H., Wilkoff B.L. et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017; 14 (12): e503–e551. doi: 10.1016/j.hrthm.2017.09.001.

Supplementary files

Supplementary Files
Action
1. JATS XML
2. Рис. 1. Отчёт первичного опроса электрокардиостимулятора пациента. Слева данные о динамике порога и импеданса желудочкового и предсердного электродов в течение года, справа — о ежемесячной динамике импеданса желудочкового электрода

Download (23KB)
3. Рис. 2. Данные рентгеноскопии. Электроды проведены через добавочную левую верхнюю полую вену; А — предсердный электрод; V — желудочковый электрод, красными стрелками1 обозначен перелом желудочкового электрода

Download (13KB)
4. Рис. 3. Данные рентгеновской компьютерной томографии; А — предсердный электрод; V — желудочковый электрод; красными стрелками обозначен перелом желудочкового электрода

Download (30KB)

© 2020 Kalinin R.E., Suchkov I.A., Povarov V.O., Shitov I.I., Potehinskiy S.S., Solovov D.V., Peshkov S.A.

Creative Commons License

This work is licensed
under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.





Согласие на обработку персональных данных с помощью сервиса «Яндекс.Метрика»

1. Я (далее – «Пользователь» или «Субъект персональных данных»), осуществляя использование сайта https://journals.rcsi.science/ (далее – «Сайт»), подтверждая свою полную дееспособность даю согласие на обработку персональных данных с использованием средств автоматизации Оператору - федеральному государственному бюджетному учреждению «Российский центр научной информации» (РЦНИ), далее – «Оператор», расположенному по адресу: 119991, г. Москва, Ленинский просп., д.32А, со следующими условиями.

2. Категории обрабатываемых данных: файлы «cookies» (куки-файлы). Файлы «cookie» – это небольшой текстовый файл, который веб-сервер может хранить в браузере Пользователя. Данные файлы веб-сервер загружает на устройство Пользователя при посещении им Сайта. При каждом следующем посещении Пользователем Сайта «cookie» файлы отправляются на Сайт Оператора. Данные файлы позволяют Сайту распознавать устройство Пользователя. Содержимое такого файла может как относиться, так и не относиться к персональным данным, в зависимости от того, содержит ли такой файл персональные данные или содержит обезличенные технические данные.

3. Цель обработки персональных данных: анализ пользовательской активности с помощью сервиса «Яндекс.Метрика».

4. Категории субъектов персональных данных: все Пользователи Сайта, которые дали согласие на обработку файлов «cookie».

5. Способы обработки: сбор, запись, систематизация, накопление, хранение, уточнение (обновление, изменение), извлечение, использование, передача (доступ, предоставление), блокирование, удаление, уничтожение персональных данных.

6. Срок обработки и хранения: до получения от Субъекта персональных данных требования о прекращении обработки/отзыва согласия.

7. Способ отзыва: заявление об отзыве в письменном виде путём его направления на адрес электронной почты Оператора: info@rcsi.science или путем письменного обращения по юридическому адресу: 119991, г. Москва, Ленинский просп., д.32А

8. Субъект персональных данных вправе запретить своему оборудованию прием этих данных или ограничить прием этих данных. При отказе от получения таких данных или при ограничении приема данных некоторые функции Сайта могут работать некорректно. Субъект персональных данных обязуется сам настроить свое оборудование таким способом, чтобы оно обеспечивало адекватный его желаниям режим работы и уровень защиты данных файлов «cookie», Оператор не предоставляет технологических и правовых консультаций на темы подобного характера.

9. Порядок уничтожения персональных данных при достижении цели их обработки или при наступлении иных законных оснований определяется Оператором в соответствии с законодательством Российской Федерации.

10. Я согласен/согласна квалифицировать в качестве своей простой электронной подписи под настоящим Согласием и под Политикой обработки персональных данных выполнение мною следующего действия на сайте: https://journals.rcsi.science/ нажатие мною на интерфейсе с текстом: «Сайт использует сервис «Яндекс.Метрика» (который использует файлы «cookie») на элемент с текстом «Принять и продолжить».