Mathematical simulation of a membrane capacitive compensation pressure transducer


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

The operation of a membrane capacitive pressure transducer with electrostatic compensation of the membrane deformation has been mathematically simulated. It is shown that the theory describing these instruments is incorrect and the error in measuring pressure by them may be as high as 25–29% of the measured value, which makes them inappropriate for use in pressure metrology.