A Fluorescence Spectrum Analyzer Based on a Fiber-Optic Y-Circulator


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

We consider a method of probe analysis employing fiber-optic a composite Y-circulator design for the implementation of laser-induced fluorescence (LIF) spectroscopy diagnostics. The proposed principle of the LIF analyzer is based on spatially separated directed optical beams of excitation and fluorescence radiations. Conditions of the effective Y-circulator operation as a passive fiber-optic switch are determined. Experimental investigation of the switched radiation losses is performed. An example of using the proposed analyzer for LIF diagnostics of biological media is presented.

About the authors

E. K. Algazinov

Voronezh State University

Email: shulgin@cs.vsu.ru
Russian Federation, Voronezh, 394006

V. A. Shul’gin

Voronezh State University

Author for correspondence.
Email: shulgin@cs.vsu.ru
Russian Federation, Voronezh, 394006

I. A. Lavrinenko

Voronezh State University

Email: shulgin@cs.vsu.ru
Russian Federation, Voronezh, 394006

A. A. Sirota

Voronezh State University

Email: shulgin@cs.vsu.ru
Russian Federation, Voronezh, 394006

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2018 Pleiades Publishing, Ltd.