Influence of Inclination Angle on the Polyethylene Insert Wear in Experimental Module of Hip Endoprosthesis

Cover Page

Cite item

Full Text

Abstract

Localization and the area of polyethylene insert wear in friction pair ofthe experimental module of hip joint (HJ) endoprosthesis was studied under laboratory dynamic conditions simulating its biomechanics at HJ external rotation contracture. It was stated that localization and the area of wear depended on the angle of horizontal inclination infriction pair of HJ endoprosthesis. Almost the same areas of polyethylene wear in anterior and posterior insert segments were observed when inclination angle was equal to that of the hip joint. Reduction of inclination angle resulted in the increase of wear surface in the anterior segment and its decrease in the posterior one. Displacement of the wear zone into the anterior segment was accompanied by the reduction of total insert wear area. At repeated loadings it created conditions for the higher polyethylene insert wear.

About the authors

A. I Kolesnik

Kursk State Medical University Russia

Email: ko-lesnik@mail.ru
доктор мед. наук, профессор кафедры хирургических болезней ФПО Курского ГМУ; Тел.: +7 (960) 678-90-32 30500, Курск, ул. К. Маркса, д. 3

N. S Gavryushenko

N.N. Priorov Central Institute of Traumatology and Orthopaedics

доктор техн. наук, профессор, рук. испытательной лаборатории ЦИТО им. Н.Н. Приорова

V. G Bulgakov

N.N. Priorov Central Institute of Traumatology and Orthopaedics

канд. биол. наук, старший науч. сотр. ОЭТО ЦИТО им. Н.Н. Приорова

E. B Frolov

Детская ОКБ

- врач травматолог-ортопед детской ОКБ г. Белгорода Белгород, РФ

L. V Fomin

N.N. Priorov Central Institute of Traumatology and Orthopaedics

инженер-метролог испытательной лаборатории ЦИТО им. Н.Н. Приорова

References

  1. Солодилов И.М., Колесник А.И., Сизых С.Г., Алпеев А.В., Гонеев С.В., Кравченко С.А. и др. Результаты эндопротезирования больных коксартрозом с наружной ротационной контрактурой тазобедренного сустава. Курский научно-практический вестник «Человек и его здоровье». 2012; 2: 117-22.
  2. Тихилов Р.М., Шубняков И.И., ред. Руководство по хирургии тазобедренного сустава. т. 1. СПб: РНИИТО им. Р.Р. Вредена; 2014.
  3. Загородний Н.В., Каграманов С.В., Чрагян Г.А. Ревизионное эндопротезирование тазобедренного сустава бесцементными компонентами эндопротеза Цваймюллера (Bicon, SL и SLR-plus). Вестник травматологии и ортопедии им. Н. Н. Приорова. 2012; 1: 66-72.
  4. Кавалерский Г.М., Мурылев В.Ю., Петров Н.В., Силин Л.Л., Рукин Я.А. Асептическое расшатывание эндопротеза тазобедренного сустава. М.: Медицина; 2011.
  5. Жабин А.И., Носкова Т.И., Гаврюшенко Н.С., Захарова В.И., Образцова Л.Н., Курзин А.А. Протезы тазобедренных суставов. Лабораторная оценка степени износа трущихся поверхностей. Основные рекомендации. Р 42-610-02. М.: ВНИИИМТ; 2002.
  6. Колесник А.И., Гонеев С.В, Соколенко Н.В., Гаврюшенко Н.С., Андросов В.В., Гончаров Н.Н. и др. Способ экспресс-моделирования износа полиэтиленового вкладыша металлической чашки или полиэтиленовой чашки в динамических условиях при разных углах горизонтальной инклинации в экспериментальном модуле эндопротеза тазобедренного сустава. Патент РФ № 2550730; 2015.
  7. Колесник А.И., Гонеев С.В, Гаврюшенко Н.С., Соколенко Н.В., Андросов В.В., Гончаров Н.Н. и др. Экспериментальный модуль эндопротеза тазобедренного сустава для экспресс-моделирования износа полиэтиленового вкладыша металлической чашки или полиэтиленовой чашки эндопротеза тазобедренного сустава. Полезная модель РФ № 153973; 2015.
  8. Докалин А.Ю., Фролов Е.Б., Чеботарева Т.М., Сизых С.Г., Колесник А.И., Соколенко Н.В. и др. Клинико-биомеханическое соответствие наружной ротационной контрактуры и горизонтальной инклинации экспериментального модуля тазобедренного сустава. В кн.: Материалы межрегиональной научно-практической конференции «Актуальные вопросы травматологии, ортопедии и комбустиологии». Воронеж: 2014: 24-.
  9. Ахтямов И.Ф., Кузин В.В., Кузьмин И.И., Рыков Ф.Г., Туренков С.В. Новые способы профилактики интраоперационных осложнений эндопротезирования тазобедренного сустава. Гений ортопедии. 2007; 2: 71-5.
  10. Скипенко Т.О. Отдаленные результаты тотального эндопротезирования тазобедренного сустава с керамика-керамической парой трения: Автореф. дис. … канд. мед. наук. М.; 2014.
  11. Kadar T., Furnes O., Aamodt A., Indrekvam K., Havelin L., Haugan K., Espehaug B., Hallan G. The influence of acetabular inclination angle on the penetration of polyethylene and migration of the acetabular component: a prospective, radiostereometric study on cemented acetabular components. J. Bone Joint. Surg. Br. 2012; 94 (3): 302-7.
  12. Johansson H.R., Johnson A.J., Zywiel M.G., Naughton M., Mont M.A., Bonutti P.M. Does acetabular inclination angle affect survivorship of alumina-ceramic articulations? Сlin. Orthop. Relat. Res. 2011; 469 (6): 1560-6.
  13. Hart A.J., Buddhdev P., Winship P., Faria N., Powell J.J., Skinner J.A. Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing. Hip. Int. 2008; 18 (3): 212-9.
  14. Wan Z., Boutary M., Dorr L.D. The influence of acetabular component posi tion on wear in total hip arthroplasty. J. Arthroplasty. 2008; 23 (1): 51-6.
  15. Patil S., Bergula A., Chen P.C., Colwell C.W. Jr, D'Lima D.D. Polyethylene wear and acetabular component orientation. J. Bone Joint. Surg. Am. 2003; 85 (Suppl. 4): 56-63.
  16. Little N.J., Busch C.A., Gallagher J.A., Rorabeck C.H., Bourne R.B. Acetabular polyethylene wear and acetabular inclination and femoral offset. Clin. Orthop. Relat. Res. 2009; 467 (11): 2895-2900.
  17. Korduba L.A., Essner A., Pivec R., Lancin P., Mont M.A., Wang A., Delanois R.E. Effect of acetabular cup abduction angle on wear of ultrahigh-molecular-weight polyethylene in hip simulator testing. Am. J. Orthop. (Belle Mead NJ). 2014; 43 (10): 466-71.
  18. Košak R., Kralj-Iglič V., Iglič A., Daniel M. Polyethylene wear is related to patient-specific contact stress in THA. Clin. Orthop. Relat. Res. 2011; 469 (12): 3415-22.
  19. Rijavec B., Košak R., Daniel M., Kralj-Iglič V., Dolinar D. Effect of cup inclination on predicted contact stress-induced volumetric wear in total hip replacement. Comput. Methods Biomech. Biomed. Engin. 2015; 18 (13): 1468-73.

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2016 Eco-Vector



Согласие на обработку персональных данных с помощью сервиса «Яндекс.Метрика»

1. Я (далее – «Пользователь» или «Субъект персональных данных»), осуществляя использование сайта https://journals.rcsi.science/ (далее – «Сайт»), подтверждая свою полную дееспособность даю согласие на обработку персональных данных с использованием средств автоматизации Оператору - федеральному государственному бюджетному учреждению «Российский центр научной информации» (РЦНИ), далее – «Оператор», расположенному по адресу: 119991, г. Москва, Ленинский просп., д.32А, со следующими условиями.

2. Категории обрабатываемых данных: файлы «cookies» (куки-файлы). Файлы «cookie» – это небольшой текстовый файл, который веб-сервер может хранить в браузере Пользователя. Данные файлы веб-сервер загружает на устройство Пользователя при посещении им Сайта. При каждом следующем посещении Пользователем Сайта «cookie» файлы отправляются на Сайт Оператора. Данные файлы позволяют Сайту распознавать устройство Пользователя. Содержимое такого файла может как относиться, так и не относиться к персональным данным, в зависимости от того, содержит ли такой файл персональные данные или содержит обезличенные технические данные.

3. Цель обработки персональных данных: анализ пользовательской активности с помощью сервиса «Яндекс.Метрика».

4. Категории субъектов персональных данных: все Пользователи Сайта, которые дали согласие на обработку файлов «cookie».

5. Способы обработки: сбор, запись, систематизация, накопление, хранение, уточнение (обновление, изменение), извлечение, использование, передача (доступ, предоставление), блокирование, удаление, уничтожение персональных данных.

6. Срок обработки и хранения: до получения от Субъекта персональных данных требования о прекращении обработки/отзыва согласия.

7. Способ отзыва: заявление об отзыве в письменном виде путём его направления на адрес электронной почты Оператора: info@rcsi.science или путем письменного обращения по юридическому адресу: 119991, г. Москва, Ленинский просп., д.32А

8. Субъект персональных данных вправе запретить своему оборудованию прием этих данных или ограничить прием этих данных. При отказе от получения таких данных или при ограничении приема данных некоторые функции Сайта могут работать некорректно. Субъект персональных данных обязуется сам настроить свое оборудование таким способом, чтобы оно обеспечивало адекватный его желаниям режим работы и уровень защиты данных файлов «cookie», Оператор не предоставляет технологических и правовых консультаций на темы подобного характера.

9. Порядок уничтожения персональных данных при достижении цели их обработки или при наступлении иных законных оснований определяется Оператором в соответствии с законодательством Российской Федерации.

10. Я согласен/согласна квалифицировать в качестве своей простой электронной подписи под настоящим Согласием и под Политикой обработки персональных данных выполнение мною следующего действия на сайте: https://journals.rcsi.science/ нажатие мною на интерфейсе с текстом: «Сайт использует сервис «Яндекс.Метрика» (который использует файлы «cookie») на элемент с текстом «Принять и продолжить».