Measurement of the Spectral Irradiance of Deuterium Lamps in the Range of Wavelengths 200–400 nm Using a High-Temperature Black Body Model
- Авторлар: Khlevnoy B.B.1, Solodilov M.V.1
-
Мекемелер:
- All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
- Шығарылым: Том 61, № 11 (2019)
- Беттер: 1098-1105
- Бөлім: Optophysical Measurements
- URL: https://bakhtiniada.ru/0543-1972/article/view/246639
- DOI: https://doi.org/10.1007/s11018-019-01555-8
- ID: 246639
Дәйексөз келтіру
Аннотация
A procedure is presented for measurements of the spectral irradiance of deuterium lamps in the wavelength range 200–400 nm with unit reproduction using a high-temperature blackbody model. It is shown that the expanded measurement uncertainty varies between 1.6–6.9%. The measurement results were confirmed with calibrations between VNIIOFI (Russia) and PTB (Germany). The divergence of the results obtained by the two laboratories does not exceed 2%.