Estimating the Error Component Associated with the Selection of a Gas Flow Model in Critical Nozzles


Дәйексөз келтіру

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

A more accurate model has been developed for an ideal compressed gas flow in a critical nozzle used as a standard for volumetric flow rate. An error assessment is carried out for the inadequacy of the classical model of a critical nozzle that is applied to obtaining the actual value of the volumetric flow rate produced by a nozzle in operating conditions. It is shown that the error of model inadequacy must be considered to be in the upper bounds region of the effective measurement ranges of gas counters when calibrating them by means of volumetric flow rate standards on critical nozzles.

Авторлар туралы

V. Chesnokov

All-Russia Research Institute of Metrological Service (VNIIMS)

Хат алмасуға жауапты Автор.
Email: chesnokov@vniims.ru
Ресей, Moscow

B. Ipolitov

All-Russia Research Institute of Metrological Service (VNIIMS)

Email: chesnokov@vniims.ru
Ресей, Moscow

Қосымша файлдар

Қосымша файлдар
Әрекет
1. JATS XML

© Springer Science+Business Media, LLC, 2017