详细

The possibility of double spectral conversion of UV radiation from the spectral range of 240–280 nm to the visible spectral region using luminescent glasses and fibers with antimony ions and molecular silver clusters is shown. Planar and fiber prototypes of UV radiation dosimeters are tested. It is shown that the sensitivities of dosimeter of the first and second types are 1400 and 360 mV/W/m2, respectively.