The Development of a Multiparametric Immunochromatographic Test System for the Analysis of Cardiomarkers


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

Abstract—An immunochromatographic test system for the combined analysis of two cardiomarkers, such as fatty acid binding protein (FABP) and troponin I (TnI), in serum and whole blood is developed. The main stages of the design of a multiparametric system are investigated, including the influence of the antibody position and gold nanoparticle size on the results. The detection limits of the method are found to be 10 ng/mL for FABP and 1 ng/mL for TnI, and the coefficient of variation did not exceed 10%. The developed test system allows us to confirm the diagnosis of acute myocardial infraction (N = 9). The analysis time is 15 min.

作者简介

I. Andreeva

Department of Chemistry, Moscow State University

Email: aposipov@mail.ru
俄罗斯联邦, Moscow

E. Yakovleva

Department of Chemistry, Moscow State University

Email: aposipov@mail.ru
俄罗斯联邦, Moscow

V. Grigorenko

Department of Chemistry, Moscow State University

Email: aposipov@mail.ru
俄罗斯联邦, Moscow

A. Osipov

Department of Chemistry, Moscow State University; National Research Technological University MISiS

编辑信件的主要联系方式.
Email: aposipov@mail.ru
俄罗斯联邦, Moscow; Moscow

补充文件

附件文件
动作
1. JATS XML

版权所有 © Allerton Press, Inc., 2019