TXRF determination of mercury(II) in water in combination with liquid–liquid microextraction
- Authors: Oskolok K.V.1, Monogarova O.V.1, Alov N.V.1
-
Affiliations:
- Department of Chemistry
- Issue: Vol 72, No 4 (2017)
- Pages: 174-177
- Section: Article
- URL: https://bakhtiniada.ru/0027-1314/article/view/163461
- DOI: https://doi.org/10.3103/S002713141704006X
- ID: 163461
Cite item
Abstract
A highly selective hybrid way of TXRF determination of mercury(II) in drinking water at the level of n(10–2–100) μg/L is developed. The technique of preconcentration of mercury(II) ions includes directly suspended droplet microextraction with benzene in the form of an iodide molecular complex. The proposed method of determination is characterized by its high degree of sensitivity and reproducibility (cmin = 8 ng/L, sr = 0.12 (100 ng/L)). The accuracy of the analysis results is confirmed by the introduced–found method.
About the authors
K. V. Oskolok
Department of Chemistry
Author for correspondence.
Email: k_oskolok@mail.ru
Russian Federation, Moscow, 119991
O. V. Monogarova
Department of Chemistry
Email: k_oskolok@mail.ru
Russian Federation, Moscow, 119991
N. V. Alov
Department of Chemistry
Email: k_oskolok@mail.ru
Russian Federation, Moscow, 119991
Supplementary files