A Prototype of a Ring Neutron Detector
- Авторлар: Baskov V.A.1, Govorkov B.B.1, Poliansky V.V.1
-
Мекемелер:
- Lebedev Physical Institute, Russian Academy of Sciences
- Шығарылым: Том 62, № 3 (2019)
- Беттер: 305-308
- Бөлім: Nuclear Experimental Technique
- URL: https://bakhtiniada.ru/0020-4412/article/view/160691
- DOI: https://doi.org/10.1134/S0020441219030035
- ID: 160691
Дәйексөз келтіру
Аннотация
When the half-ring of a prototype neutron detector with an inner radius of 50 mm is calibrated using a 60Co ionizing-radiation source, its relative detection efficiency gradually decreases from ~15% at the center of the half-ring to ~5% at its ends. When calibrated using cosmic muons, the relative detection efficiency is approximately 100% over the entire detector volume. The presence of zones with reduced and increased efficiency values are associated with the intricate detector geometry.