A Method for Increasing the Sensitivity of Mass-Spectrometric Measurements


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

A method for increasing the sensitivity of mass-spectrometric measurements in the analysis of small amounts of gases, which are insufficient for organizing a stationary flow through an ion source, is considered. Measurements are carried out upon a single injection of a portion of a gas. The use of this method for increasing the sensitivity in isotope helium analyses on MI9303 magnetic-resonance mass spectrometers allowed the sensitivity of the spectrometer to be increased to ~105 atoms in a sample, i.e., it allowed measurements for a helium volume in a sample of up to ~10−14 cm3. The experience of carrying out isotope analyses in samples of inert gases of various origins using different methods for evacuating chambers of mass analyzers was generalized.