Energy Broadening Model for BGO Array Crystal from 0.059 to 1.332 MeV


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

The energy resolution and energy broadening parameters are of great importance as far as scintillator detector is concerned. At present, BGO crystal due to its high density, high yield of photoelectric absorption, availability and cost is used in particle physics and medical imaging system especially in positron emission tomography (PET). The purpose of this study is to calculate the Gaussian energy broadening parameters of a 10 × 10 BGO array in order to use it in simulation for analysis of its spectral response in the presence of gamma rays in a wide range of energies. The Gaussian energy broadening parameters were extracted from the experimental spectrum to use them as the input parameters in MCNP4c code. The simulated response of the detector employing these parameters showed a good agreement with the experimental data.

Sobre autores

Mojtaba Askari

Nuclear Science and Technology Research Institute

Autor responsável pela correspondência
Email: m_askari_l@yahoo.com
Irã, Tehran

Ali Taheri

Nuclear Science and Technology Research Institute

Email: m_askari_l@yahoo.com
Irã, Tehran

Mohammad Sasanpour

Nuclear Science and Technology Research Institute

Email: m_askari_l@yahoo.com
Irã, Tehran

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Pleiades Publishing, Inc., 2018