Highly-sensitive analog magnetometer based on a null-picovoltmeter
- Авторы: Chiang Y.N.1, Dzyuba M.O.1
-
Учреждения:
- Verkin Institute for Low Temperature Physics and Engineering
- Выпуск: Том 59, № 4 (2016)
- Страницы: 565-568
- Раздел: General Experimental Techniques
- URL: https://bakhtiniada.ru/0020-4412/article/view/159195
- DOI: https://doi.org/10.1134/S0020441216030234
- ID: 159195
Цитировать
Аннотация
The principle of measurement of small magnetic moments using superconducting null-picovoltmeter (SNPV) is described. It is shown that in the mode with optimal voltage sensitivity (≥10–12 V), the sensitivity of SNPV to a magnetic moment can reach 10–8 emu, the level only implemented in SQUIDs. An example is given of using SNPV to measure the magnetic susceptibility of superconducting diamagnetic and paramagnetic samples, approximately 1 mm3 in volume, with different signs of the magnetization in the magnetizing alternating magnetic fields with induction of about 1 G, at low temperatures.
Об авторах
Yu. Chiang
Verkin Institute for Low Temperature Physics and Engineering
Автор, ответственный за переписку.
Email: chiang@ilt.kharkov.ua
Украина, pr. Lenina 47, Kharkov, 61103
M. Dzyuba
Verkin Institute for Low Temperature Physics and Engineering
Email: chiang@ilt.kharkov.ua
Украина, pr. Lenina 47, Kharkov, 61103