Measurement of the Dead Time of an Optoelectronic Streak Camera with Different Linear Sweep Ranges


Дәйексөз келтіру

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

A method for measuring the dead time of an optoelectronic streak camera with different linear sweep ranges is described. The proposed method was used for the PN-1 streak camera. Photographs of the time-resolved emission of a DRSh350-2 mercury arc flash lamp for linear sweep ranges of 50 and 100 ns are presented. The dead times of the PN-1 streak camera for linear sweep ranges of 25, 50, and 100 ns were 5, 75, and 105 ns, respectively. The photographs of time-resolved second-harmonic spectra emitted by plasmas of copper and aluminum targets that were obtained in experiments on the interaction of nanosecond laser radiation with the matter using a streak camera are presented.

Авторлар туралы

T. Abbasov

Lebedev Physical Institute

Email: artsahakyan@yandex.ru
Ресей, Moscow, 119991

A. Kalinenkov

Lebedev Physical Institute

Email: artsahakyan@yandex.ru
Ресей, Moscow, 119991

V. Puzyrev

Lebedev Physical Institute

Email: artsahakyan@yandex.ru
Ресей, Moscow, 119991

A. Sahakyan

Lebedev Physical Institute

Хат алмасуға жауапты Автор.
Email: artsahakyan@yandex.ru
Ресей, Moscow, 119991

A. Starodub

Lebedev Physical Institute

Email: artsahakyan@yandex.ru
Ресей, Moscow, 119991

B. Fakhriev

Lebedev Physical Institute

Email: artsahakyan@yandex.ru
Ресей, Moscow, 119991

O. Yakusheva

Lebedev Physical Institute

Email: artsahakyan@yandex.ru
Ресей, Moscow, 119991

Қосымша файлдар

Қосымша файлдар
Әрекет
1. JATS XML

© Pleiades Publishing, Inc., 2018