A Calorimeter Based on Peltier Modules
- 作者: Volkov A.S.1, Tyagunin A.V.1, Koposov G.D.1
-
隶属关系:
- Northern (Arctic) Federal University named after M.V. Lomonosov
- 期: 卷 61, 编号 5 (2018)
- 页面: 766-768
- 栏目: Laboratory Techniques
- URL: https://bakhtiniada.ru/0020-4412/article/view/160435
- DOI: https://doi.org/10.1134/S0020441218040267
- ID: 160435
如何引用文章
详细
A two-chamber calorimeter with industrial TEC1-12706 Peltier modules used as heat-flow sensors and heating elements is described. The measured temperature dependence of the volt–watt sensitivity in the temperature range from –180 to 50°С is presented. The features of the measuring technique are described. The measured temperature dependence of the specific heat of ice is shown.