Comparison of an Avalanche Photodiode and a Photomultiplier Tube as Photodetectors of Near-Infrared Radiation in the Photon-Counting Mode
- 作者: Nadeev A.I.1, Razenkov I.A.1, Shelefontyuk D.I.1, Shevtsov E.S.1
-
隶属关系:
- Institute of Atmosphere Optics, Siberian Branch
- 期: 卷 61, 编号 3 (2018)
- 页面: 377-381
- 栏目: General Experimental Techniques
- URL: https://bakhtiniada.ru/0020-4412/article/view/160224
- DOI: https://doi.org/10.1134/S0020441218020057
- ID: 160224
如何引用文章
详细
The results of testing near-infrared photodetectors, that is, a HAMAMATSU H10330B-25 photomultiplier tube (PMT) and a receiving module that was created by the authors on the basis of an SPAD PGA-284 avalanche photodiode, are presented. The results show that when weak optical signals are registered with an avalanche-photodiode-based detector, its sensitivity is higher than that of the PMT by one order of magnitude, while its maximum counting frequency is 10 times lower.
作者简介
A. Nadeev
Institute of Atmosphere Optics, Siberian Branch
编辑信件的主要联系方式.
Email: nadeev@iao.ru
俄罗斯联邦, Tomsk, 634021
I. Razenkov
Institute of Atmosphere Optics, Siberian Branch
Email: nadeev@iao.ru
俄罗斯联邦, Tomsk, 634021
D. Shelefontyuk
Institute of Atmosphere Optics, Siberian Branch
Email: nadeev@iao.ru
俄罗斯联邦, Tomsk, 634021
E. Shevtsov
Institute of Atmosphere Optics, Siberian Branch
Email: nadeev@iao.ru
俄罗斯联邦, Tomsk, 634021
补充文件