Mass-Spectrometric Observation of C+ Ions during Electrospray with In-Source Atomization


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

Carbon C+ ions formed in a mass-spectrometric electrospray source with controlled in-source fragmentation and atomization interface have been observed for the first time. The measurements were performed in a special MI-20 LowMass mass spectrometer (MS-Bio LLC) using aqueous methanol solutions of lithium and beryllium salts. The peak of C+ ions was substantially broadened as compared to the peaks of metal ions. A model is proposed that explains the formation of C+ ions due to charge exchange immediately at the high-vacuum boundary near the skimmer output.

Sobre autores

A. D’yachenko

Institute for Analytical Instrumentation, Russian Academy of Sciences

Email: kolomna.88@mail.ru
Rússia, St. Petersburg, 198095

N. Blashenkov

Ioffe Physical Technical Institute, Russian Academy of Sciences

Email: kolomna.88@mail.ru
Rússia, St. Petersburg, 194021

N. Samsonova

Ioffe Physical Technical Institute, Russian Academy of Sciences

Autor responsável pela correspondência
Email: kolomna.88@mail.ru
Rússia, St. Petersburg, 194021

L. Gall

Institute for Analytical Instrumentation, Russian Academy of Sciences

Email: kolomna.88@mail.ru
Rússia, St. Petersburg, 198095

A. Semenov

A.A. Bochvar All-Russia Advanced Research Institute for Inorganic Materials

Email: kolomna.88@mail.ru
Rússia, Moscow, 123098

A. Lizunov

A.A. Bochvar All-Russia Advanced Research Institute for Inorganic Materials

Email: kolomna.88@mail.ru
Rússia, Moscow, 123098

N. Gall

Ioffe Physical Technical Institute, Russian Academy of Sciences

Email: kolomna.88@mail.ru
Rússia, St. Petersburg, 194021

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Pleiades Publishing, Ltd., 2019