Optimization of N-acetylcysteine Determination by Ion Mobility Spectrometry using Central Composite Design


Цитировать

Полный текст

Открытый доступ Открытый доступ
Доступ закрыт Доступ предоставлен
Доступ закрыт Только для подписчиков

Аннотация

In this study, a simple and rapid method for N-acetylcysteine (NAC) determination was proposed using positive corona discharge ion mobility spectrometry. To optimize cell and injection temperature, central composite design was used. Other parameters such as corona and drift voltage, carrier and drift gas flow rate, and pulse width were adjusted. The calibration curve was linear under optimum conditions in the range of 3–100 ng with a correlation coefficient (R2) of 0.99. The relative standard deviations were 8.1 and 6.1%, for 20 and 50 ng of analyte (n = 9), respectively. Limits of detection and quantification were 1 and 3 ng, respectively. The developed method was successfully applied for the determination of NAC in pharmaceutical and human plasma samples.

Об авторах

L. Tehrani

Department of Chemistry, Yazd Branch, Islamic Azad University

Email: sheibani@iauyazd.ac.ir
Иран, Yazd

A. Sheibani

Department of Chemistry, Yazd Branch, Islamic Azad University

Автор, ответственный за переписку.
Email: sheibani@iauyazd.ac.ir
Иран, Yazd

A. Nazari

Department of Art & Architectural, Yazd Branch, Islamic Azad University

Email: sheibani@iauyazd.ac.ir
Иран, Yazd

Дополнительные файлы

Доп. файлы
Действие
1. JATS XML

© Pleiades Publishing, Ltd., 2018

Согласие на обработку персональных данных

 

Используя сайт https://journals.rcsi.science, я (далее – «Пользователь» или «Субъект персональных данных») даю согласие на обработку персональных данных на этом сайте (текст Согласия) и на обработку персональных данных с помощью сервиса «Яндекс.Метрика» (текст Согласия).