GET 198–2017, State Primary Standard of Unit of Power of Magnetic Losses and Unit of Magnetic Induction of Constant Magnetic Field in Range from 0.1 to 2.5 T and Magnetic Flux from 1·10–5 to 3·10–2 Wb


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

Information to improve GET 198–2011, the State Primary Standard of the unit of power of magnetic losses, is presented. A plant for reproduction of the unit of magnetic induction of a constant magnetic field in the range from 0.1 to 2.5 T and unit of magnetic flux in the range from 1·10–5 to 3·10–2 Wb is introduced as a standard. Studies to improve the standard were carried out in 2015–2016. The composition, operating principle, and metrological characteristics of GET 198–2017, the State Primary Standard of the unit of power of magnetic losses and unit of magnetic induction of a constant magnetic field in the range from 0.1 to 2.5 T and of magnetic flux in the range from 1·10–5 to 3·10–2 Wb, are described. Information on preliminary comparisons is presented.

作者简介

E. Volegova

Ural Research Institute of Metrology; Ural Federal University

编辑信件的主要联系方式.
Email: ekaterina.volegova@gmail.com
俄罗斯联邦, Ekaterinburg; Ekaterinburg

M. Malygin

Ural Research Institute of Metrology

Email: ekaterina.volegova@gmail.com
俄罗斯联邦, Ekaterinburg

T. Maslova

Ural Research Institute of Metrology

Email: ekaterina.volegova@gmail.com
俄罗斯联邦, Ekaterinburg

A. Volegov

Ural Research Institute of Metrology; Ural Federal University

Email: ekaterina.volegova@gmail.com
俄罗斯联邦, Ekaterinburg; Ekaterinburg

补充文件

附件文件
动作
1. JATS XML

版权所有 © Springer Science+Business Media, LLC, part of Springer Nature, 2018