Preparation of Calibration Gas Mixtures by the Chromato-Desorption Method for Increasing the Accuracy of Quantitative Determination of Biogenic Pentane in Expired Air


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

The possibility of increasing the accuracy of preparing n-pentane calibration gas mixtures by using injection-type chromato-desorption microsystems is investigated. The possibility of obtaining calibration gas mixtures in the range of 4–12 mg/m3max = 16%) is shown. It is established that the change in the composition of the mixture during storage for three months is not more than 7%.