Metrological Support of Liquid and Aerosol Particle Counters


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

The calibration and verification of liquid and aerosol particle counters are described. Aerosol particle counters are calibrated using latex spheres of known size. Depending on the design of the counters, the light scattered by the particles is detected at different angles. It is shown that detection of the scattered light at an angle of 90° leads to large measurement errors if the refractive index of the detected particles is not specified a priori. When the scattered light is detected over the entire solid angle, the error is substantially reduced. The metrological characteristics of counters with different designs are analyzed.

作者简介

E. Lesnikov

All-Russia Research Institute of Physicotechnical and Radio Measurements (VNIIFTRI)

编辑信件的主要联系方式.
Email: lesnikov@vniiftri.ru
俄罗斯联邦, Mendeleevo, Moscow Region

D. Balakhanov

All-Russia Research Institute of Physicotechnical and Radio Measurements (VNIIFTRI)

Email: lesnikov@vniiftri.ru
俄罗斯联邦, Mendeleevo, Moscow Region

V. Dobrovol’skii

All-Russia Research Institute of Physicotechnical and Radio Measurements (VNIIFTRI)

Email: lesnikov@vniiftri.ru
俄罗斯联邦, Mendeleevo, Moscow Region

补充文件

附件文件
动作
1. JATS XML

版权所有 © Springer Science+Business Media New York, 2017