Use of the Four-Point Method for Measuring the Homogeneity of the Coating Deposited Inside Ceramic Vacuum Chambers of Pulsed Magnets for the MAX-IV Synchrotron Radiation Facility
- Авторы: Anchugov O.V.1, Shvedov D.A.1
-
Учреждения:
- Budker Institute of Nuclear Physics, Siberian Branch, Russian Academy of Sciences
- Выпуск: Том 62, № 3 (2019)
- Страницы: 440-443
- Раздел: Laboratory Techniques
- URL: https://bakhtiniada.ru/0020-4412/article/view/160773
- DOI: https://doi.org/10.1134/S0020441219030023
- ID: 160773
Цитировать
Аннотация
An instrument has been developed and manufactured to measure the uniformity of the layer thickness deposited inside ceramic vacuum chambers of electron-beam injection devices (kickers). Using this instrument, it is possible to measure this parameter at an aperture size starting with 1 cm and at different aperture geometries. The measurement range is 0.1−100 µm. The necessity of such measurements is substantiated, various types of sensors are described, and results for some vacuum chambers of the injection kickers in the MAX-IV synchrotron radiation facility (Sweden) are presented.