Comparison of an Avalanche Photodiode and a Photomultiplier Tube as Photodetectors of Near-Infrared Radiation in the Photon-Counting Mode
- Авторлар: Nadeev A.I.1, Razenkov I.A.1, Shelefontyuk D.I.1, Shevtsov E.S.1
-
Мекемелер:
- Institute of Atmosphere Optics, Siberian Branch
- Шығарылым: Том 61, № 3 (2018)
- Беттер: 377-381
- Бөлім: General Experimental Techniques
- URL: https://bakhtiniada.ru/0020-4412/article/view/160224
- DOI: https://doi.org/10.1134/S0020441218020057
- ID: 160224
Дәйексөз келтіру
Аннотация
The results of testing near-infrared photodetectors, that is, a HAMAMATSU H10330B-25 photomultiplier tube (PMT) and a receiving module that was created by the authors on the basis of an SPAD PGA-284 avalanche photodiode, are presented. The results show that when weak optical signals are registered with an avalanche-photodiode-based detector, its sensitivity is higher than that of the PMT by one order of magnitude, while its maximum counting frequency is 10 times lower.